8.0 F
& P
USES
OWER
8.1
Fuses
F1
Internal
supply fuses
F2
Auxiliary
circuit fuses
F3
Heat Light
fuses
Customer
fuses
Table of internal fuse and external supply ratings.
The phases are generally unequal on these models. A figure equal to or exceeding the highest
phase current is shown, and the wiring and fusing should match or exceed this figure. The function
of the customer's (external) fuse is to protect the customer's wiring.
Model
GHA, GVA 12/300
as above, 2-ph+N
GHA, GVA 12/450
as above, 3-phase
GHA, GVA 12/600
GHA, GVA 12/750
GHA, GVA 12/900
GHA, GVA 12/1050
GHA, GVA 12/1200
GHC, GVC 12/450
as above, 3-phase
GHC, GVC 12/600
GHC, GVC 12/750
GHC, GVC 12/900
GHC, GVC 12/1050
GHC, GVC 12/1200
GHC, GVC 12/900E
GHC, GVC 12/1050E
GHC, GVC 12/1200E
GHC, GVC 12/1500
MF48 -1.18
S
ETTINGS
F1-F3: Refer to the circuit diagrams.
Fitted to all G range models.
For "universal" wiring each circuit has 2 fuses, in
most cases fitted on board to the EMC filter.
Fitted to all G range models.
In most cases fitted on board to the EMC filter.
Not fitted in standard G range models (wire size
does not reduce).
Required to protect customer supply circuit and
lead.
internal
supply
fuses
2 x 12.5A
1 x 12.5A
2 x 16A
6 x 7A
6 x 10A
6 x 10A
6 x 10A
6 x 12.5A
6 x 12.5A
2 x 16A
6 x 7A
6 x 7A
6 x 12.5A
6 x 12.5A
6 x 12.5A
6 x 12.5A
6 x 10
6 x 12.5A
6 x 16A
6 x 25A
external:
external:
3-phase+N
1-phase
380/220-
200-240V
415/240V
12A
--
--
6A/ph
15A
--
--
6A/ph
20A
10A/ph
25A
10A/ph
30A
10A/ph
30A
12A/ph
35A
12A/ph
15A
--
--
6A/ph
20A
10A/ph
25A
12A/ph
30A
12A/ph
30A
12A/ph
35A
12A/ph
30A
10A/ph
30A
12A/ph
35A
15A/ph
43A
21A/ph
G range
on-board and up to 16 Amps:
32mm x 6mm type F
2 Amps
On board: 20mm x 5mm type F
Other: 32mm x 6mm type F
if fitted:
32mm x 6mm type F
See rating label for amperage;
see table below for minimum
supply and fuse rating.
external:
3-phase
delta
200-240V
--
--
--
10A/ph
12A/ph
15A/ph
20A/ph
20A/ph
20A/ph
--
10A/ph
12A/ph
15A/ph
20A/ph
20A/ph
20A/ph
20A/ph
20A/ph
25A/ph
25A/ph
19