Troubleshooting
Please refer to the table below to identify problems that persist with no messages appearing.
PROBLEM
CAUSED BY
The meter does not
Batteries are dead or there is a
power on
problem with the meter.
Test does not start
Poor amount of sample
after applying blood
sample
Problem with the test strip
The result is doubtful
There may be a problem with the
test strip
The codes of the meter and of the
test strip vial do not match
[ 27 ]
finetest
TM
WHAT TO DO
Change the batteries and if the problem persists, contact
your local representative.
Apply a sufficient amount of the blood sample to a new test
strip and re-test.
Re-test with a new strip.
Re-test with a new strip.
Re-code the meter and re-test.
Specifications
Specifications
Model
IGM-0005A
Test Range
0.6 ~ 33.3 mmol/L (10 ~ 600 mg/dL)
Reading Time
9 sec
Memory Capacity
200 test results
Operating Temperature
50 ~ 86˚F(10 ~ 40˚C )
Operating Humidity
10 ~ 90%
Sample Type
Capillary Whole Blood
Calibration
Plasma-Equivalent
Hematocrit
30 ~ 55%
Sample Volume
Minimum 1.5 µl
Display Type
LCD
Size (D X W X H)
81
56
23 (mm)
X
X
Weight
Less than 40g ± 1(including batteries)
Power Source
3V Li Battery (CR2032)
1
X
Battery Life
running 5000 tests
Glucose Unit
Either mmol/L or mg/dL
INFOPIA
[ 28 ]