Troubleshooting
MESSAGE
CAUSED BY
No saved data
The ambient temperature is
too low
The ambient temperature is
too high
[ 25 ]
finetest
TM
WHAT TO DO
Place the meter at the operating temperature range for more
than 10 minutes, and re-test.
Place the meter at the operating temperature range for more
than 10 minutes, and re-test.
Troubleshooting
MESSAGE
CAUSED BY
The test result is lower than
10mg/dL (0.6mmol/L)
The test result is higher than
600mg/dL (33.3mmol/L)
WHAT TO DO
If you believe that your result is incorrect, recheck the meter by using
the finetest
TM
Control Solution. With normal result, re-test with blood
samples two or three times. " LO" results indicate severe
hypoglycemia (low blood glucose). If the "LO"message persists,
please contact your physician or diabetes healthcare professional.
If you believe that your result is incorrect, recheck the meter by using
the finetest
Control Solution. With normal result, re-test with blood
TM
samples two or three times. " HI" results indicate severe
hyperglycemia (high blood glucose). If the "HI"message persists,
please seek medical attention immediately.
INFOPIA
[ 26 ]
Need help?
Do you have a question about the finetest IGM-0005A and is the answer not in the manual?
Questions and answers
Здравствуйте хотел бы узнать какая погрешность данного глюкометра если она вообще