Specifications; Troubleshooting - Infopia Element Lite Instructions For Use Manual

Blood glucose monitoring system
Table of Contents

Advertisement

Troubleshooting

Odpravljanje težav
Message
Problem
The test result is
higher than
600mg/dL
Battery is dead or
The blood glucose
test meter does not
there is a problem
power on
with
the
glucose test meter
Test does not start
Poor blood sample
after applying blood
sample
There may be a problem
The result is
with the blood glucose
inconsistent
test strip
No readings in
memory to display
Not enough
readings in
memory to display
designated
averages
38
www.infopia21.com
Action Required
Check the accuracy of your
blood glucose test strip by
performing a glucose control
solution test. If the test results in
a normal reading, re-test your
blood sample two or three
times. If "Hi" persists, consult
your doctor immediately.
Change the battery and if the
problem persists, please contact
your local representative.
blood
Please insert a new blood glucose
test strip and perform your test
again(refer to the P.23)
Please insert a new blood glucose
test strip and
perform your test
again

Specifications

Sample type
Sample volume
Test Range
Reading time
Calibration
Hematocrit
Altitude
Operating Temperature
Operating Humidity
Test Strip Storage
Display Type
Dimension
Weight
Power rating
Battery Life
Capillary whole blood
minimum 0.3㎕
10~600 mg/dL (0.55~33.31mmol/L)
3 seconds
Plasma - Equivalent
20~60%
3048 meter Up to (10,000 feet)
10 ~ 40℃ (50 ~ 104℉)
10-90%
Store 2~30℃(36~86℉)and no direct
sunlight.Do not Freeze
LCD
84.5 X 52 X 17.5±1 (mm)
41±1g ( Including Battery)
3 V Li Battery (CR2032) X 2
One year after purchasing