NOTE

DC Adjustment Procedure

Description
NOTE
64 Adjustments
The use of the preceding commands is described in the DC Adjustment
Procedure and the AC Flatness Adjustment Procedure in this chapter.
CALibration:DATA[:DC] <
constants to the AFG. It is recommended that the DC calibration
procedure in this chapter be performed instead of using this
command. The query form returns the current DC constants in
IEEE-488.2 definite block data format.
CALibration:DATA:AC <
constants to the AFG. It is recommended that the AC calibration
procedure in this chapter be performed instead of using this
command. The query form returns the current constants in
IEEE-488.2 definite block data format.
CALibration:STATe:AC <
using the calibration constants. If
used. If
state
is OFF, corrections will not be used.
<
>
A DC adjustment is performed on the AFG by reading a series of voltages
output by the AFG, then entering those values back into the AFG. After all
measurements have been completed, new calibration constants are
calculated and stored in non-volatile memory. To ensure accuracy, perform
the DC calibration procedure at one-year intervals.
This procedure uses a firmware routine to adjust the AFG's DC calibration
constants. The
CALibration[:DC]:BEGin
sequence and the
CALibration[:DC]:POINt?
next calibration point. All AFG settings are performed by the firmware
routine.
If an error occurs at any time during the procedure, abort (
Firmware revision A.01.02 or later is required for this procedure. Use the
command to find out which firmware revision you are using.
*IDN?
block
transfers the DC calibration
>
block
transfers the AC flatness calibration
>
state
enables or disables AC corrections
>
state
is ON, corrections will be
<
>
command starts the DC calibration
command steps the AFG to the
*RST
) and start over.
Chapter 3