100 Verification Tests - C Programs
iprintf (id, "FREQ 1.0E7;\n");
iprintf (id, ":VOLT 10DBM\n");
iprintf (id, "INIT:IMM\n");
/*--------------------Perform Test-----------------------------------------*/
for (i = 0;i <= 8;i ++)
{
printf ("\n\n Set Spectrum Analyzer Start Frequency to %.1e Hz", start[i]);
printf ("\n Set Spectrum Analyzer Stop frequency to %.1e Hz", stop[i]);
printf ("\n Measure amplitude of the highest peak");
printf ("\n Enter amplitude (in dBm) : ");
scanf ("%lf", &peak);
printf ("\n\nResult = %.2f", peak-10);
printf ("\n\nPress ENTER to continue");
getchar ();
}
iprintf (id, "*RST;*CLS\n");
iclose (id);
}
/* Close instrument session */
Appendix A