HP E1340A Service Manual page 29

Function generator
Hide thumbs Also See for E1340A:
Table of Contents

Advertisement

Chapter 2
Perform steps 3 - 4 for each amplitude listed in Table 2-1:
3. Set the AFG output amplitude:
<amplitude>
VOLT
where <amplitude> is the value specified in Table 2-1.
4. Trigger the DMM and record the reading in Table 2-7.
Table 2-1. DC Accuracy Test Points
Amplitude
Specifications
(volts)
10.235 ± 0.276
10.235
5.120 ± 0.276
5.120
2.560 ± 0.276
2.560
0.320 ± 0.276
0.320
0.000 ± 0.276
0.000
-0.320 ± 0.276
-0.320
-2.560 ± 0.276
-2.560
-5.120 ± 0.276
-5.120
-10.240 ± 0.276
-10.240
(volts)
Verification Tests 31