HP E1340A Service Manual page 53

Function generator
Hide thumbs Also See for E1340A:
Table of Contents

Advertisement

Chapter 2
2. Set the AFG to output a 10 dBm, 10 MHz sinewave:
FREQ 1.0E7;
:VOLT 10DBM
INIT:IMM
Perform steps 3 and 4 for each frequency range listed in Table 2-10:
3. Set the spectrum analyzer start frequency and stop frequency to the
values listed in Table 2-6.
4. Measure the amplitude (in dBm) of the highest peak. Subtract the
amplitude of the fundamental (10 dBm) from the reading and record
the result:
result ( dBc) = reading (dBm) − (10 dBm)
Table 2-6. Spurious/Non-Harmonic Test Points
Start
Frequency
Frequency
(Hz)
1 E6
10.5 E6
21 E6
31 E6
41 E6
51 E6
75 E6
100 E6
100 E6
125 E6
125 E6
150 E6
Set AFG frequency
Set AFG amplitude
Initiate waveform
Stop
Spec
(Hz)
(dBc)
9.5 E6
-43
19 E6
-43
29 E6
-43
39 E6
-43
49 E6
-43
75 E6
-43
-43
-43
-43
Verification Tests 55