Perform steps 3 - 4 for each amplitude and filter listed in Table 2-3:
3. Set the AFG output amplitude:
<amplitude>
VOLT
where <amplitude> is the value specified in Table 2-3.
4. Trigger the DMM and record the reading in Table 2-7.
Table 2-3. AC Accuracy Test Points
Specifications
Amplitude
Low Limit
(volts)
(volts)
7.239
6.889
6.450
6.138
5.750
5.472
4.566
4.345
2.881
2.742
1.440
1.370
0.229
0.218
High Limit
(volts)
7.606
6.777
6.042
4.798
3.027
1.513
0.241
Verification Tests 37