Equipment Setup
Adjustment Procedure
Chapter 3
Perform a complete autocalibration on the DMM (unless an
–
autocal has been performed within the last 24 hours)
Connect the equipment as shown in Figure 3-1
–
Set the DMM to DCV, 1 V range, 10 NPLC
–
Figure 3-1. DC Adjustment Setup
1. Reset the AFG:
*RST;*CLS
2. Send the command to start the DC adjustment routine and wait for
the command to complete:
CAL:DC:BEGIN
*OPC?
The AFG will return a "1" when ready.
Reset AFG and clear status registers
Initiate DC cal
Adjustments 65