Test 2-1: DC Accuracy
Description
Equipment Setup
Test Procedure
30 Verification Tests
The purpose of this test is to verify that the AFG meets its specifications for
DC accuracy.
Connect equipment as shown in Figure 2-3
–
Set DMM to DCV, autorange
–
Figure 2-3. DC Accuracy Test Setup
1. Reset the AFG:
*RST;*CLS
2. Set up the AFG to output a DC signal:
FUNC DC;
:OUTP:LOAD INF
Reset AFG and clear status registers
Select DC waveform
Infinite load
Chapter 2