6.10 Adjusting the adjustment weight

In the case of KERN ABS/ABJ, the external adjustment weight can be displayed
variably. Capture the value of the adjustment weight here.
The selected adjustment weight must be used during the adjustment.
Menu call:
SEttinG
CAL dEF
CAL SEt
Sequence:
(1)
Press and hold the CAL/MENU key until the function SettinG appears on the
display. Press the TARE/
(2)
Press and hold the CAL/MENU key until CAL SEt appears on the display.
Press the TARE/
(3)
Use the PRINT/ key
(4)
Use the UNIT/ key
(5)
If the weight has changed, press the TARE/
Use the ON/OFF/ESC key to return to weighing mode.
ABS/ABJ-BA-e-1019
- key
- key
to select the figure you wish to change (from left to right).
to increase the numeric value of the selected figure
- key to check the value weight.