5.4
Checking the balance with the internal weight (ABJ only)
(1)
Remove the items being weighed from the balance plate and press the
TARE/
key; the display will indicate zero.
-
(2)
Press the CAL/MENU key until FUnC.SEL appears. Then press the TARE/
key.
(3)
When CAL appears, press the TARE/
(4)
Press the CAL/MENU key until itESt appears, then press the TARE/
check will start.
(5)
E tESt 2 appears. The zero point is being checked.
(6)
E tESt 1 appears. The internal weight is being checked.
(7)
E tESt 0 appears. The zero point is being checked again.
(8)
The check is finished. The difference to the previous adjustment will be
displayed.
(9)
If no error message is issued, the balance displays tEStEND for some seconds
and then it will display zero again, thereby indicating that the check has been
successful.
28