Counting - KERN ABS Operating Instruction

Analytical balance
Table of Contents

Advertisement

6.6

Counting

Display symbol: PCS
The counting program allows the weight calculation to be converted into number of
parts with respect to a reference weight. The displayed weight will be taken for a preset
number of parts (setting for 10, 20, 50 or 100 parts).
Counting accuracy is higher when there are more parts. The minimum reference weight
is 0.01 g.
Condition:
Change the units to PCS,
see Chapter 6.4 "Changing the units"
Sequence:
(1)
Remove the items being weighed from the balance plate and press the
TARE/
- key; the display will indicate zero.
(2)
Place the parts to be weighed on the plate
(3)
Use the UNIT/ key
(4)
Press and hold the CAL/MENU key (2 sec) until SEt 10 appears on the display.
(5)
Use the CAL/MENU key to select the respective number of parts, for the parts
located on the balance plate (SEt 10 PCS, SEt 20PCS, SEt 50 PCS,
SEt 100 PCS).
(6)
Press the TARE/
may now gradually add more parts to the balance plate; the display is expressed
in number of parts.
Note:
If the "Err 20" error message is displayed, the reference weight is too small for the
number of parts.
ABS/ABJ-BA-e-1019
to select the change of units in PCS
-key to display the number of parts that has selected. You
33