Starting the balance

2.5.1 Warm-up time

2.5.2 On and off (standby mode)

2.5.3 Auto-check

oFF
Standby
CHE 5
12
In order for the balance to provide accurate results, the
device must be warmed up for at least 4 hours after being
plugged for the first time or after an extended power
outage. After this time, the balance will have reached the
necessary operating temperature.
To turn the display on and off, use the ON/OFF/ESC key.
An automatic function test of the balance electronics is
carried out after switching on from standby mode. At the
end of the test, the display will indicate zero. The balance is
now ready to be used.
The following symbols will appear in the display window of
the balance for information purposes:
The balance is in OFF mode.
The balance is not plugged into the power mains.
(New connection, power outage above 1 second)
The balance is in standby mode.
The display was turned off by pressing the ON/OFF/ESC
key and is now available. The balance is ready to run when
turned on and does not require a warm-up period. In
addition, the time is displayed on the ABJ model in standby
mode.
The CHE display appears after the power is turned on.
The balance will work through a check procedure
and then switch to OFF mode.