Setting The Time (Abj Only); Entering Seconds - KERN ABS Operating Instruction

Analytical balance
Table of Contents

Advertisement

6.13 Setting the time (ABJ only)

6.13.1 Entering seconds

Menu call:
SEttinG
CAL dEF
t-HH-MM
SEC AdJ
HH-MM-SS
Sequence:
(1)
Press and hold the CAL/MENU key until the function setting appears. Press the
TARE/
- key.
(2)
Press and hold the CAL/MENU key until t-HH-MM appears (HH:hour, MM:min.)
(3)
Press the TARE/
(4)
Press the TARE/
between 00 and 29 this will be rounded down to zero sec; if pressed between 30
and 59 this will be rounded up to the next full minute.
(5)
Use the ON/OFF/ESC key to return to weighing mode. The setting is stored.
40
- key, SEC AdJ will appear.
- key to set the desired time. If the TARE/
- key is pressed
ABS/ABJ-BA-e-1019