Ac Current Verification - Keithley DMM6500 Calibration Manual

6 1/2-digit multimeter with scanning
Hide thumbs Also See for DMM6500:
Table of Contents

Advertisement

Section 2: Performance verification
Verify digitize current 100 mA range
Description
UUT zero
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify digitize current 1 A range
Description
UUT zero
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify digitize current 3 A range
Description
UUT zero
⅔ scale (+)
Half scale (+)
Half scale (–)
⅔ scale (–)
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy
limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
Verify digitize current 10 A range
See
Digitize current 10 A range verification
37).

AC current verification

The following topics describe how to verify AC current.
Verify AC current on the 100 µA to 3 A ranges
To verify AC current accuracy, you will:
Apply accurate voltages from the Fluke 5720A or 5730A multifunction calibrator to the DMM6500
front-panel terminals.
Verify that the displayed readings fall within specified limits.
2-30
Model DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Calibrator setpoint
Lower limit
(A)
0
REL
0.1
9.99200E-02
0.05
4.99450E-02
–0.05
–5.00550E-02
–0.1
–1.00080E-01
Calibrator setpoint
Lower limit
(A)
0
REL
1
9.99900E-01
0.5
4.99350E-01
–0.5
–5.00650E-01
–1
–1.00100E+00
Calibrator setpoint
Lower limit
(A)*
0
REL
2
1.99700E+00
1.5
1.49745E+00
–1.5
–1.50255E+00
–2
–2.00300E+00
(on page 2-39) under
Upper limit
n/a
1.00080E-01
5.00550E-02
–4.99450E-02
–9.99200E-02
Upper limit
n/a
1.00100E+00
5.00650E-01
–4.99350E-01
–9.99900E-01
Upper limit
n/a
2.00300E+00
1.50255E+00
–1.49745E+00
–1.99700E+00
Rear-panel verification
DMM6500-905-01 Rev. A/July 2018
(on page 2-