OPERATION
How To Operate
4 Press power switch on level
1 Set the instrument to the tripod or smooth
5 Select the precision mode by
surface.
2 Make sure instrument is roughly level.
3 Press power switch (ON).
Height Alert Function
Example Operational
When the instrument system detects a shock, this function informs
the operator of it.
• When the instrument's installation status (height) is sharply changed
by the contact of the operator or the like, this function stops auto
leveling to keep the operation accuracy and informs the operator of the
situation. The three lamps blink at the same time as shown at the right.
• After 1 minute has passed since the auto leveling function was
activated and the laser beam was emitted, this function works.
• The height alert function does not work in the "Manual" mode.
Aligning Direction of Slope
It is possible to set slope only in the X axis direction (refer to

"CHECKS AND ADJUSTMENTS").

[1. Installation of the instrument]
When using the laser to set the slope, the laser must be
properly aligned so the slope axis of the laser beam is parallel
to the desired direction of slope.
The sight on top of the instrument is calibrated to the slope
axis of the laser beam. Follow the steps below to align the
laser to the desired direction of slope:
1 Establish a target line parallel to desired direction of slope.
2 Set up the laser over this line (drop a plumb bob from the
tripod mounting screw).
3 Rough align the instrument to the direction of slope. Make
sure it is properly oriented for the slope to be set.
4 Using the sight, position the instrument so the sight is
centered on the target. (see illustration at right)
Target
CHECKS AND ADJUSTMENTS
1 Checking and adjusting calibration
Horizontal calibration of the laser beam can be checked by the user.
[Checking]
1 Set up a tripod approx. 50m (160ft) from a wall. Mount the instrument on the
tripod, facing the X1 side toward the wall.
2 Turn the instrument on and allow auto-leveling to complete.
3 Put the level sensor in fine detection mode by pressing the On-Grade precision
switch.
4 By using the level sensor, mark the center position of laser beam on the wall. (X1)
5 Turn off the instrument.
Loosen the tripod screw, rotate the instrument 180 degrees and re-secure it on
the tripod. The X2 side of the instrument faces toward the wall.
When rotating the instrument, avoid changing the height.
6 Turn the unit on again and allow auto-leveling to complete.
7 By using the level sensor, mark the center position of laser beam on the wall. (X2)
8 If the difference value of marked two laser beam heights (difference value of X1
and X2) are less than 5mm, adjustments are not needed.
The difference value is greater than 5mm, adjust the instrument as described in
right.*
9 Check the Y side as the same way.
* If the difference value is greater than 40mm (±90"), it exceeds the
adjustment range. Please contact your Topcon dealer.
2 Checking cone error
Cone error
Datum
Wall A
Wall B Wall A
position

ERROR CODE

Use the table below to determine operation errors indicated by blinking lamps on the control panel.
If corrective action listed does not correct error, please contact your local Topcon dealer.
Lamp Indication
Error Code
Lamp B, C and D blink in turn
Auto-leveling range error
Lamp A lights
Battery power error
Lamp B, C and D blink
Height alert error
simultaneously
Lamp D blinks quickly
Calibration error
Lamp E (Red) lights
Checking and adjusting
Lamp B, C and D blink in turn
calibration error
Lamp A, B, C and D blink
Internal error
simultaneously
REGULATION
FCC
NOTE:
This equipment has been tested and found to comply with limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interfer-
ence to radio communications.
However, there is no guarantee that interference will not occur in a particular instal-
lation.
If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encour-
aged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio / TV technician for help.
6 Locate the on-grade position "---" by moving
the level sensor up and down.
sensor (ON).
7 Mark the position of On-Grade index.
(Top of the level sensor is 40mm [1 9/16"]
pressing the On-Grade
from index for offset marking.)
precision switch.
Shock is given to the instrument.
[2. Setting slope]
Operating procedure
(Setting slope)
1 Press Power switch ON.
You can set the slope after auto-leveling is completed.
2 Press one of Slope keys once. The laser beam keeps slop-
ing in the direction of the pressed key.
3 Press the Slope key again. The laser beam stops sloping.
(Unless you press the Slope key again, the laser beam is set
horizontally.)
4 Press the Slope key to adjust the slope. How long you press
the Slope key can change the laser beam sloping speed.
The blinking speed of the lamp provides a visual indication of
the beam movement speed.
5 Press Power switch OFF to cancel the slope setting.
How to set the slope before the power is OFF again
1 To set the slope before the power is OFF again, turn on the Power switch
as pressing either right or left of the Slope keys. The slope is kept and you
can adjust the slope in the same way as Step 4.
(When the Power switch is ON without the Slope key pressed, the laser
beam will return to level.)
X2 Laser beam
Approx. 50m
X2
Handle
Control panel
Approx. 40m
Wall B
(For the lamp indication, refer to "Lamp position".)
Corrective Action
Correct tilt of the instrument until it less than 3 degrees.
Replace the four alkaline dry cell batteries with new ones at a time or charge
the battery pack.
Turn power off, rough level the instrument, then turn power on again. Check
height of laser beam as it may have changed.
Repeat calibration procedure. If error repeats contact your local Topcon dealer.
Push the slope key of the opposite side, and align slope.
Turn power off, then on again. If error repeats contact your local Topcon dealer.
Declaration of Conformity
Model Number:RL-H4C
Trade Name:TOPCON CORPORATION
Manufacturer
Name:TOPCON CORPORATION
Address:75-1, Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580 JAPAN
Country:JAPAN
U.S.A. Representative
Responsible party:TOPCON POSITIONING SYSTEMS,INC.
Address:7400 National Drive Livermore, CA94551, U.S.A
Telephone number:925-245-8300
LS-80L Indicator
Height alert warning of rotating laser*1
A flash and a buzzer sound signifies
that the height alert function of rotating
laser is operating.
Rotating laser battery warning*2
A flash shows that the rotating laser
power is low.
Note: The warning displays *1 and *2 are the functions that the
level sensor detects alarm signal from the rotating laser.
The level sensor can be canceled the alarm detection from the
rotating laser.
To be canceled the detection; Press the power switch while
pressing the buzzer sound switch when powering on.
Height alert status
The three lamps blink at the
same time and the rotary head
rotates at low speed.
Key
operation
Power switch
During auto-leveling
After auto-leveling is completed. : Lamp B (Green) lights.
After auto-leveling
is completed.
Lamp E (Green) blinks quickly.
Slope key
Slope key
Lamp E (Green) lights.
Slope key
When Slope key is not pressed : Lamp E (Green) lights.
Slope key
Press for a shorter time : Blinks slowly
Press for a longer time : Blinks quickly
Power switch
All lamps are OFF.
• When you set the "Manual" mode to ON after setting the slope, the beam is fixed at the set slope position.
At this time the Slope key does not function.
For information about Manual mode ON/OFF, refer to "NOMENCLATURE AND FUNCTIONS".
• Check the beam often during slope use for slope accuracy. Check instrument calibration periodically.
Laser point of X1
[To calibrate the X axis]
1 Face the X1 side of the instrument toward a wall, press the Power switch while pressing the height alert OFF key. Then the
Wall
height alert OFF lamp will light, and manual mode ON lamp will blink. (X axis is selected.)
level sensor
2 Press the height alert OFF key to calibrate the X axis. The manual mode ON lamp will light. When auto-leveling finishes, the
laser beam will emit.
3 Using the level sensor, mark the on-grade height of laser beam on a wall.
X1 Laser beam
X1
4 Rotate the instrument 180 degrees to face X2 side toward a wall.
5 In the same way as step 3, mark the on-grade height of laser beam on a wall.
6 Press the slope key to make adjustment so that the laser beam height may be at the center between the positions of Step 3
and Step 5.
Approx. 50m
7 Press the height alert OFF key to memorize the new laser beam calibration. The height alert OFF lamp will blink. Power will
shut off automatically when the calibration memorization is complete.
The X axis adjustment is completed now.
X1
[To calibrate the Y axis]
1 Face the Y1 side of the instrument (Control panel side) toward a wall, press the Power switch while pressing the height alert
OFF key. Then the height alert OFF lamp will light, and manual mode ON lamp will blink. (X axis is selected.)
2 Press the right Slope key once again. The auto leveling lamp will light. (Y axis is selected.)
X2
3 Press the height alert OFF key to calibrate the Y axis. The auto leveling lamp will light.
4 Using the level sensor, mark the on-grade height of laser beam on a wall.
5 Rotate the instrument 180 degrees to face Y2 side toward a wall.
6 In the same way as step 4, mark the on-grade height of laser beam on a wall.
7 Press the slope key to make adjustment so that the laser beam height may be at the center between the positions of Step 4
and Step 6.
Y2
8 Press the height alert OFF key to memorize the new laser beam calibration. The height alert OFF lamp will blink. Power will
shut off automatically when the calibration memorization is complete.
The Y axis adjustment is completed now.
X1
If the height alert OFF lamp blinks quickly and the power is not automatically turned off when pressing the
height alert OFF key to memorize the height, the height exceeds the adjustment range. Please contact
your local Topcon dealer.
Y1
To discontinue calibration the instrument, press the Power switch.
Perform the following check after completing horizontal calibration procedure.
[Checking]
1 Set up the laser centered between two walls approximately 40m (131ft) apart. Orient the instrument so one axis, either X or Y, is facing the walls.
2 Locate and mark the position of the rotating laser beam on both walls using the level sensor.
3 Turn off the instrument and move the instrument closer to wall A (1m to 2m /3 ft to 6 ft). Do not change the axis orientation of the instrument. Turn the
instrument on.
4 Again locate and mark the position of the rotating laser beam on both walls using the level sensor.
5 Measure the distance between the first and second marks on each wall. If the difference between each set of marks is less than 4mm (5/32 of an inch),
no error exists.
* If the difference value is greater than 4mm (5/32 inch), contact your Topcon dealer.

SPECIFICATIONS

RL-H4C
Laser source
Laser output
Safety standard for laser beam : CDRH (FDA) Class IIIa, IEC Class 3R
Automatic correction range
Accuracy
Manual slope settable range
Rotational speeds
Operating range
Power supply/Operating time
Charging time
Operating time
Protection against water and dust
Operating temperature
Storable temperature range
LS warning display
Dimensions
Laser beam height
Weight (Dry battery type)
(Rechargeable battery type) : 2.6kg [5.7 lbs] (Including BT-74Q)
Tripod screw
LS-80L
Beam detection window
Beam detection precision
Beam detection indication
Power source
Operating time
Auto shut-off delay
Protection against water and dust
Operating temperature
Storage temperature
Dimensions
Weight
High precision mode
Normal precision mode
Higher than datum position
(Buzzer sound: High frequent beep sound)
Move the sensor downward.
Datum position
(Buzzer sound: Continuous beep sound)
Lower than datum position
(Buzzer sound: Low frequent beep sound)
Move the sensor upward.
Battery remaining display
(1)
(2)
(3)
[How to reset]
1 Turn off the power switch.
2 Check whether the instrument is installed correctly.
3 Turn on the power switch. Auto leveling starts again. After
auto leveling is finished, the laser beam is emitted.
4 Make sure that the laser beam is set at the correct height.
Then, restart the operation.
Lamp display
(Refer to the right illustration)
: Lamp B (Green) blinks.
A: Battery power
B: Power lamp /
Auto leveling lamp
C: Manual mode ON
Lamp E (green) Beam movement
D: Height alert OFF
: Moves slowly
: Moves quickly
: Laser diode (Visible, 635nm)
: 2.4mW
: ±5°
: ±10" (1mm/20m)
: ±5° (When the instrument is installed on the 0° surface)
The slope range is increased or decreased according to the tilt of the surface on which the
instrument is installed.
: 600r.p.m
: Diameter Approx. 2m to 800m (When using LS-80L)
: 4 x D size dry cell batteries (alkaline)*1or Ni-MH battery pack BT-74Q (7000mAh)
: Approx. 13 hours (Using with AD-15)
: Approx. 100 hours (Using with alkaline manganese drybattery / at +20°C (+68°F))
: Approx. 60 hours (Using with Ni-MH battery pack BT-74Q/ at +20°C (+68°F))
: IP66 (Based on thestandard IEC60529)
: -20°C to +50°C (-4°F to +122°F)
: -30°C to +60°C (-22°F to +144°F)
: RL-H4C height alert warning(Warning is displayed on the indicator of LS-80L.)
RL-H4C battery warning (Warning is displayed on the indicator of LS-80L.)
: 177(L)x196(W)x217(H)mm (7.0 x 7.7 x 8.5 in)
: 187mm (Height from the instrument's bottom surface to the center point of laser beam)
: 2.4kg [5.3 lbs] (Including dry cell batteries)
: 5"/8X11 threads for surveying instrument
: 50mm (2.0 in)
: High precision: ±1mm (±0.04 in)/Normal precision: ±2mm (±0.08 in)
: Liquid crystal (both sides) and buzzer
: 2 x AA size dry cell batteries
: Approx. 120 hours at +20°C (+68°F) (Using alkaline manganese dry cell batteries)
: Approx. 30 minutes without beamdetection
: IP66 (Based on thestandard IEC60529)
: -20°C to +50°C (-4°F to +122°F)
: -30°C to +60°C (-22°F to +140°F)
: 146(L)x76(W)x26(H)mm (5.7x2.9x1.0 in)
: 0.19 kg [0.41 lbs] (including dry cell batteries)
http://www.topcon.co.jp
Detective Range
(1) Battery is sufficient.
(2) The power is low, but
laser is still usable.
(3) Dead battery.
Replace the dry battery
with new one.
Lamp Position
E: Slope lamp
Slope key
lamp
lamp
lamp
Height alert
Manual mode
Power switch
OFF key
ON key
Please see the attached address list or
the following website for contact addresses.
31365 90031
GLOBAL GATEWAY http://global.topcon.com/