Error Message; Instrument Error Code - Topcon IM-600 Instruction Manual

Digital illuminance meter
Table of Contents

Advertisement

5. ERROR MESSAGE

5.1

INSTRUMENT ERROR CODE

Error code
Explanation
E1
Zero
adequately.
E2
The measured value exceeds measurable
range.
E3
Improper calculation.
1.
2.
E4
The value affected by correction factor
exceeds the display range 999900 lx.
E5
Measured
exceeds 1,000,000,000 lx・h.
E6
Integration time exceeds 9999 hours.
E7
Reference values, distance data, or correction
factors exceed the limit of entry.
E8
The instrument fails to write or read data.
E10
The response switch is pushed 3 times or
more during a zero adjustment.
●If an error message is still displayed even after you take above action, repair may be required.
Contact TOPCON TECHNOHOUSE or the dealer where you purchased this instrument.
46
adjustment
does
The value exceeds 9999% in the %
mode.
The value exceeds 999000cd in cd
mode.
value
of
not
complete
integral
illuminance
Action
Check the connection between
Display unit and Detector unit.
If no problem is found, contact
your dealer or
TOPCON TECHNOHOUSE.
Check the measurement range
and use upper range.
1. The
difference
measured
value
reference value is too large.
2. Check that the distance data
is correct.
Check if the correction factor is
correct.
Do not exceed the limit of integral
illuminance.
Do not exceed the limit of
integration time.
Check the entry range.
Contact your dealer or TOPCON
TECHNOHOUSE.
Don't touch the response switch
during a zero adjustment.
between
and