5. ERROR DISPLAY

5.1 INSTRUMENT ERROR CODE

This chapter will explain the error codes that are displayed on the instrument and acquired by the
"ERR" command.
Instrument code Code
0
E-01
1
E-02
2
E-03
3
E-04
4
E-05
5
52
Contents
Normal
Recognition of detector
unit has failed.
Zero calibration has
failed.
Temperature error
Integral irradiance is
exceeded.
"Over range" error
Remedial measure
Remedial measure is not necessary.
Check whether the detector unit is
connected to the control unit correctly.
Check whether the detector window is
covered against light.
Lower the internal temperature.
The upper limit of the integral irradiance
measurement is exceeded.
measurement again.
Set a measurement range properly.
Perform