5. ERROR MESSAGE

5.1

ERROR CODE FOR INSTRUMENT

Error code
Explanation
E1
Appears when the zero adjustment is not
complete.
E2
In manual range mode, appears when the
measured value exceeds the display range.
In auto range mode, appears when the
measured value exceed measurement range
at range 4.
E3
Appears when the value exceeds 9999% in
the % measurement mode.
E4
Appears when a correction factor is effective
and the obtained value exceeds the display
range 999900 μW/cm
E5
Appears when measured vale in integral
irradiance exceeds 1,000,000,000 mJ/cm
E6
Appears when integration time exceeds
999900 seconds.
E7
Appears when entered reference values and
correction factors are out of the input range.
E8
Appears when data can not be written or read
out due to defects of EEPROM.
E10
Appears when the response switch is turned 3
times (maximum applicable times) or more
when a zero adjustment is being conducted.
●If an error message is still displayed even after the above action is conducted, repair may be
required.
Contact TOPCON TECHNOHOUSE or the dealer where you purchased this instrument.
2
.
2
Action
Check if the detector connect to
Display unit correctly.
If no problems are found, contact
your dealer or
TOPCON TECHNOHOUSE.
Check the measurement range.
The
difference
between measured value and
reference value is too large.
Check if the correction factor is
correct, and measure within the
range not exceeding the display
range.
Measure within the range not
.
exceeding the integral irradiance
range.
Measure within the range not
exceeding the integration time
range.
Check the input range.
Contact your outlet dealer or
TOPCON TECHNOHOUSE.
Never turn the response switch
when a zero adjustment is being
conducted.
in
irradiance
39