5.2 Error Code in Communication

When errors occur in the instrument, which is measuring a target in combination with PC, the
instrument sends the error messages mentioned in the table below to PC.
Error code
E001
OVER-RANGE
When the brightness of the measurement target exceeds the instrument's
measurable range, this code is returned.
Take the following remedial measures and others.
- Lower the brightness of the measurement target.
- Select a smaller measuring angle.
- Insert the neutral density filter.
E002
The [MEAS./HOLD] switch is pressed on the instrument during remote
measurement and so measurement is stopped.
E004
The external synchronizing signal is not captured normally in the instrument.
Check the synchronizing signal generating device and the connection of BNC cable.
E900 ‒ 999
Internal error occurs in the instrument. Contact our company or the retailer shop
where you purchased the instrument.