4.5 Error Code

When an error occurs in the instrument when connected with a PC and performing measurement,
the following error messages will be sent to the PC.
Error Code
E003
E004
E005
E006
E007
E008
E009
E010
E011
E012
E013
E014
E015
E016
74
Displayed when there is an abnormal measurement angle.
measurement angle is set properly.
Displayed when a measurement command is sent before executing zero
adjustment. Execute zero adjustment.
Displayed when time has elapsed since the previous calibration and it is
necessary to execute calibration by TOPCON TECHNOHOUSE factory.
Displayed when the value of the correction factor is abnormal.
Displayed when the value of the area correction factor is abnormal.
Area correction limit write error.
Displayed when the value of one side of specified squared area is greater than
0.03.
Area correction limit write error.
Displayed when specified area overlap with other area in same group.
Area correction limit write error.
Displayed when the specified area does not place on the chromaticity diagram,
or when the maximum and minimum value of chromaticity area is improper.
Area correction limit write error.
Displayed when the entered data does not correctly stored in instrument.
Displayed when the single correction/direct correction dip switch setting and
communication command setting do not match. Make sure that the dip switch
setting matches to the communication command.
Displayed when zero-adjustment is abnormal. There is a possibility that the
instrument do not make perfect dark condition during zero-adjustment, so
perform zero-adjustment with the eyepiece cap attached.
Displayed when the internal shutter is abnormal. There is a possibility that it
is broken, so contact the dealer or TOPCON TECHNOHOUSE.
Displayed when average measurement is abnormal. There is a possibility that
the number of times that the instrument perform measurement are less than
necessary in average measurement mode, due to over range.
measurement range setting for the measured object, and make sure the
measurement angle is properly set.
Communication error.
The power OFF then back ON again.