BEFORE REQUESTING SERVICE

ERROR CODE

When the EZ-200 malfunctions, an error code is displayed with the message together.
Code
Initial measurement error #0
Dust detection error #1
CMOS sensor error #9
Measurement error #10
Measurement error #20
Measurement error #30
Measurement error #50
Measurement error #60
Measurement error #70
Measurement error #80
Measurement error #90
Measurement error #100
Measurement error #499
Motor error #500
Frame motor initial error
Motor error #501
Frame motor initial error
Motor error #510
Lens motor initial error
Motor error #511
Lens motor initial error
Motor error #520
Cover motor initial error
Motor error #500
Cover motor initial error
In initial measurement, the quantity of the bright spots,
which have been transmitted through the Hartmann plate,
is not normal.
In initial measurement, stain has been found on the cover
glass.
The image sensor has malfunctioned.
The bright spots, which have been transmitted through
the Hartmann plate, are not normal.
In the "AUTO" mode, the instrument has malfunctioned
while judging a single-focal lens or progressive lens.
When measuring a single-focal lens, the instrument has
malfunctioned.
When measuring a progressive lens, the progressive
zone cannot be found.
When measuring the far point unit of a progressive lens,
the instrument has malfunctioned.
When measuring the near point unit of a progressive lens,
the instrument has malfunctioned.
The prism value cannot be measured at the inputted PD
position.
When measuring the base segment unit of a bifocal lens,
the instrument has malfunctioned.
When measuring the segment unit of a bifocal lens, the
instrument has malfunctioned.
Any other measurement error except the above has
occurred.
While the frame retainer is operating, the instrument has
malfunctioned.
While the frame retainer is operating, the instrument has
malfunctioned.
While the lens retainer is operating, the instrument has
malfunctioned.
While the lens retainer is operating, the instrument has
malfunctioned.
While the shield cover is operating, the instrument has
malfunctioned.
While the shield cover is operating, the instrument has
malfunctioned.
Meaning
BEFORE REQUESTING SERVICE
41