1.2.9 Cleaning/Disinfecting

During normal use of the POC-S175 may become soiled and should,
therefore, be cleaned regularly. Agents: Green tinctured soap and Enzy-
matic detergents
Steps:
1. Wipe the POC-S175 with a clean cloth that has been moistened in the
cleaning solution.
2. Prepare agent per manufacturer's instructions or hospital protocol.
3. Wipe thoroughly with a clean cloth
Cautions
Do not immerse or rinse the POC-S175 and its periph-
erals. If you accidentally spill liquid on the device, dis-
connect the unit from the power source. Contact your
Biomed regarding the continued safety of the unit
before placing it back in operation.
Do not spray cleaning agent on the chassis.
Do not use disinfectants that contain phenol. Do not
autoclave or clean the POC-S175 or its peripherals
with strong aromatic, chlorinated, ketone, ether, or
Esther
solvents, sharp tools or abrasives. Never immerse
electrical connectors in water or other liquids.

1.3 LCD Specifications

• Display type: 17" TFT LCD.
• Resolution: 1280 x 1024
• Colors: 16.7M (8 bits/color)
• Dot size (mm): 0.264 x 0.264
• Viewing angle: 170°
• Luminance: 250 cd/m2
• Contrast ratio: 500 : 1
5