2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in «Section 2.».
The memory retains all values although date and time (and
possibly also set alarm times) must be reset – the year number
therefore flashes automatically after the batteries are replaced.
Use 2 new, long-life 1.5V, size AAA batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the instrument is not going to be used for
a prolonged period.

Using rechargeable batteries

You can also operate this instrument using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharge as
a result of low use of the instrument, even when switched off).
Always remove the rechargeable batteries if you do not intend
to use the instrument for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
BP W100

8. Error Messages

If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too weak.
weak
Re-position the cuff and repeat the
measurement.*
«ERR 2» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle tension.
Repeat the measurement, keeping your
arm still.
«ERR 3» No pressure
An adequate pressure cannot be gener-
in the cuff
ated in the cuff. A leak may have occurred.
Check that the cuff is correctly connected
and is not too loose. Replace the batteries
if necessary. Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
7
EN