9-2 Initializing The Balance
This function returns the following parameters to factory settings. Calibration data
□ Function table
□ The sample unit mass value (counting mode),
100% reference mass value (percent mode)
□ The data that is stored in the balance using the data memory function
□ External calibration weight and target weight value
□ Function switch settings
Note Be sure to calibrate the balance after initialization.
Step1
Press the
Step2
While pressing and holding the PRINT key and the SAMPLE
key, press the ON:OFF key to display p5 .
Step3
Press the
Step4
Press the
To cancel this operation, press the
Step5
Press the
Step6
With displaying
balance. The balance will automatically return to the weighing
mode.
key to turn off the display.
ON:OFF
key to display
SAMPLE
key.
PRINT
key to change
RE-ZERO
Clr
, press the PRINT key to initialize the
Clr
.
key.
CAL
.
38
With pressing
Hold
and holding
p 5
C l r
e n d
0 . 0 0
g