Operation
1
Press the
ON:OFF
2
While pressing and holding the
key. The balance displays p5 .
ON:OFF
3
Press the
PRINT
4
Set the switches using the following keys.
SAMPLE
key
RE-ZERO
key
PRINT
key
CAL
key

8-2 Initializing the Balance

This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100%
reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical data
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the
ON:OFF
2
While pressing and holding the
SAMPLE
key, press the
The balance displays
3
Press the
SAMPLE
4
Press the
PRINT
To cancel this operation, press the
5
Press the
RE-ZERO
6
Press the
PRINT
The balance will automatically return to the weighing mode.
key to turn off the display.
PRINT
key. Then the balance displays the function switches.
To select the switch to change the parameter.
To change the parameter of the switch selected.
0: To inhibit changes.
To store the new parameter and return to the weighing mode.
To cancel the operation (
mode, press the
key to turn off the display.
PRINT
ON:OFF
p5 .
key to display
key.
CAL
key.
key to initialize the balance.
key and the
SAMPLE
1: To permit changes
Clr is displayed.) To return to the weighing
CAL
key once again.
key and the
key.
Clr .
key.
35
key, press the
With these keys
Press the key
held down