Initializing the Balance

9.2.
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
External calibration weight and target mass value
Function switch settings ("9.1. Permit or Inhibit")
Note
Be sure to calibrate the balance after initialization.
1 Press the
ON:OFF
2 While pressing and holding the
key, press the
RANGE
3 Press the
RANGE
4 Press the
PRINT
To cancel this operation, press the
5 Press the
RE-ZERO
6 Press the
PRINT
The balance will automatically return to the weighing mode.
key to turn off the display.
PRINT
key to display
ON:OFF
key to display
Clr
key.
CAL
key to display
Clr go
key to initialize the balance.
key and the
.
p5
.
key.
.
24
Press and hold
Press