Invacare Pronto M91 Service Manual page 27

With surestep
Hide thumbs Also See for Pronto M91:
Table of Contents

Advertisement

6
E06
Right Park Brake
Fault
7
E07
Remote Fault
8
E08
Controller Fault
9
E09
Communications
Fault
10
E10
General Fault
11
E11
Incompatible/
incorrect Remote
Part No. 1125038
SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
00
Right Park Brake Drive-Time
Test Failed
01
Right Park Brake Output
Enabled When Wheelchair
Idle
02
Right Park Brake Output Did
not Enable When Entering
Drive Mode
03
Right Park Brake fault during
power-up testing
04
Right park brake feedback low
during drive (park brake short)
00
Local SR Fault
(CPU, EEPROM, etc.)
01
Joystick fault at the remote
02
Speed pot fault at the remote
00
Controller fault
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing fault
09
Relay fault or precharge fault
10
Bridge fault or disable all fault
11
Electronics fault: Thermistor
12
Calibration setting fault
00
Remote connection lost
01
Low communication mode
00
General fault
00
Incompatible/incorrect
Remote
27
Check the right park brake connections
and cable.
Check the communications bus,
connections and wiring. Replace the
remote.
Check connections and wiring. Replace
power module.
Check connections and wiring. Replace
Bus cable.
Check all connections and wiring. Contact
Invacare Technical Service.
Wrong type of remote connected. Ensure
the branding of the joystick matches that
of controller unit.
Pronto® M91™and M94™