Chapter 3 Features and Functions
Measurement Configuration

Autozero

When autozero is enabled (default), the multimeter internally
disconnects the input signal following each measurement, and takes a
zero reading. It then subtracts the zero reading from the preceding
reading. This prevents offset voltages present on the multimeter's input
circuitry from affecting measurement accuracy.
When autozero is disabled, the multimeter takes one zero reading and
subtracts it from all subsequent measurements. It takes a new zero
reading each time you change the function, range, or integration time.
Applies to dc voltage, dc current, and 2-wire ohms measurements only.
Autozero is enabled when you select 4-wire ohms or ratio measurements.
The autozero mode is stored in volatile memory; the multimeter
automatically enables autozero when power has been off or after a
remote interface reset.
Front-Panel Operation: The autozero mode is set indirectly when you
set the resolution.
Resolution Choices
Fast 4 Digit
*
Slow 4 Digit
Fast 5 Digit
(default)
*
Slow 5 Digit
*
Fast 6 Digit
Slow 6 Digit
* These settings configure the multimeter just as if you had
pressed the corresponding "DIGITS" keys from the front panel.
Remote Interface Operation: The
similar effect. Autozero
Autozero
issues an immediate zero measurement.
ONCE
ZERO:AUTO {OFF|ONCE|ON}
Integration Time
Autozero
0.02 PLC
Off
1 PLC
On
0.2 PLC
Off
10 PLC
On
10 PLC
On
100 PLC
On
and
OFF
ONCE
does not issue a new zero measurement.
OFF
parameters have a
59
3