Frequency And Period Characteristics - HP 34401A User Manual

Hide thumbs Also See for 34401A:
Table of Contents

Advertisement

Chapter 8 Specifications

Frequency and Period Characteristics

Frequency and Period Characteristics
Function
Range
[ 3 ]
Frequency,
100 mV
Period [ 4 ]
to
750 V
Additional Low-Frequency Errors ( % of reading )
Frequency
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 40 Hz
40 Hz – 100 Hz
100 Hz – 300 Hz
300 Hz – 1 kHz
> 1 kHz
Transfer Accuracy ( typical )
0.0005% of reading
220
Accuracy Specifications
24 Hour
[ 2 ]
Frequency
23°C 1°C
3 Hz – 5 Hz
0.10
5 Hz – 10 Hz
0.05
10 Hz – 40 Hz
0.03
40 Hz – 300 kHz
0.006
[ 4 ]
Resolution
6
1
5
1
4
1
2
2
2
0
0.12
0.12
0
0.17
0.17
0
0.2
0.2
0
0.06
0.21
0
0.03
0.21
0
0.01
0.07
0
0
0.02
Conditions:
- Within 10 minutes and
- Within 10% of initial value.
- Following a 2-hour warm-up.
- For inputs > 1 kHz and > 100 mV.
- Using 6
- Measurements are made using accepted metrology practices.
( % of reading ) [ 1 ]
90 Day
1 Year
23°C 5°C
23°C 5°C
0.10
0.10
0.05
0.05
0.03
0.03
0.01
0.01
0.5°C.
1
digit slow resolution ( 1 second gate time ).
2
Temperature
Coefficient/°C
0°C – 18°C
28°C – 55°C
0.005
0.005
0.001
0.001

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34401-90004

Table of Contents