Triggering - HP 34401A User Manual

Hide thumbs Also See for 34401A:
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference

Triggering

Triggering
See also "Triggering," starting on page 71 in chapter 3.
First-time
SCPI users,
The multimeter's triggering system allows you to generate triggers
see page 154.
either manually or automatically, take multiple readings per trigger,
and insert a delay before each reading. Normally, the multimeter will
take one reading each time it receives a trigger, but you can specify
multiple readings (up to 50,000) per trigger.
Triggering the multimeter from the remote interface is a multi-step
process that offers triggering flexibility.
First, you must configure the multimeter for the measurement by
selecting the function, range, resolution, etc.
Then, you must specify the source from which the multimeter will
accept the trigger. The multimeter will accept a software (bus) trigger
from the remote interface, a hardware trigger from the rear-panel
Ext Trig (external trigger) terminal, or an immediate internal trigger.
Then, you must make sure that the multimeter is ready to accept
a trigger from the specified trigger source (this is called the wait-for-
trigger state).
The diagram on the next page shows the multimeter's triggering system.
4
127

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34401-90004

Table of Contents