Chapter 4 Remote Interface Reference

Simplified Programming Overview

Simplified Programming Overview
You can program the multimeter to take measurements from the remote
First-time
interface using the following simple seven-step sequence.
SCPI users,
see page 154.
1. Place the multimeter in a known state (often the reset state).
2. Change the multimeter's settings to achieve the desired configuration.
3. Set-up the triggering conditions.
4. Initiate or arm the multimeter for a measurement.
5. Trigger the multimeter to make a measurement.
6. Retrieve the readings from the output buffer or internal memory.
7. Read the measured data into your bus controller.
The MEASure? and CONFigure commands provide the most straight-
forward method to program the multimeter for measurements. You can
select the measurement function, range, and resolution all in one
command. The multimeter automatically presets other measurement
parameters (ac filter, autozero, trigger count, etc.) to default values as
shown below.
MEASure? and CONFigure Preset States
Command
AC Filter (DET:BAND)
Autozero (ZERO:AUTO)
Input Resistance (INP:IMP:AUTO)
Samples per Trigger (SAMP:COUN)
Trigger Count (TRIG:COUN)
Trigger Delay (TRIG:DEL)
Trigger Source (TRIG:SOUR)
Math Function (CALCulate subsystem)
112
MEASure? and CONFigure Setting
20 Hz (medium filter)
OFF if resolution setting results in NPLC < 1;
ON if resolution setting results in NPLC 1
OFF (fixed at 10 M for all dc voltage ranges)
1 sample
1 trigger
Automatic delay
Immediate
OFF