Chapter 3 Features and Functions
Measurement Configuration
Resolution
Front-Panel Operation: Select either the slow or fast mode for each
resolution setting. The default mode is 5 digits slow.
(continued)
See also "To Set the Resolution," on page 21.
Remote Interface Operation: You can set the resolution using the
following commands.
Specify the resolution in the same units as the measurement
function, not in number of digits. For example, for dc volts, specify the
resolution in volts. For frequency, specify the resolution in hertz.
56
5: RESOLUTION
(MEAS MENU)
function
range
CONFigure:<
> {<
MEASure:<
function
>? {<
range
<
function
>:RESolution {<
CONF:VOLT:DC 10,0.001
MEAS:CURR:AC? 1,1E-6
CONF:FREQ 1 KHZ,0.1 Hz
VOLT:AC:RES 0.05
resolution
>|MIN|MAX|DEF},{<
>|MIN|MAX|DEF},{<
resolution
resolution
>|MIN|MAX}
4
1 2
digits on the 10 Vdc range
6
1 2
digits on the 1 A range
1000 Hz input, 0.1 Hz resolution
50 mV resolution on the ac function
>|MIN|MAX|DEF}
>|MIN|MAX|DEF}