Chapter 5 Error Messages
Calibration Errors
703
Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the multimeter. You must use the same security code
to unsecure the multimeter as was used to secure it, and vice versa. The
security code may contain up to 12 alphanumeric characters. The first
character must be a letter.
704
Secure code too long
A security code was received which contained more than 12 characters.
Cal aborted
705
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the multimeter.
706
Cal value out of range
The specified calibration value (CALibration:VALue) is invalid for the
present function and range.
Cal signal measurement out of range
707
The specified calibration value (CALibration:VALue) does not match
the signal applied to the multimeter.
708
Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the
required input frequency for calibration.
No cal for this function or range
709
You cannot perform calibrations for ac current, period, continuity, diode,
ratio, or on the 100 M range.
710
Full scale correction out of range
720
Cal DCV offset out of range
721
Cal DCI offset out of range
Cal RES offset out of range
722
Cal FRES offset out of range
723
724
Extended resistance self cal failed
5
181