Fluke 719PRO User Manual page 29

Pressure calibrator
Hide thumbs Also See for 719PRO:
Table of Contents

Advertisement

Valve
Figure 3. Pump Connection
Initial Setup and Basic Pressure Generation
3.
Close the vent knob.
4.
Push the pump key and see the pressure (or
vacuum) increase until the necessary pressure is
reached.
The pump motor speed will start slowly when
pressure is low (<15 psi) for better control at low
pressures.
5.
Use the fine adjustment vernier to fine tune the
pressure/vacuum indication as necessary.
6.
To fully decrease or bleed off the pressure, slowly
turn the vent knob to the open position. When this is
done carefully, the pressure-bleed rate can be
carefully controlled and this helps when taking down-
scale pressure readings.
hix009.eps
Pressure Calibrator
Note
19