Fluke 719PRO User Manual page 68

Pressure calibrator
Hide thumbs Also See for 719PRO:
Table of Contents

Advertisement

719PRO
Users Manual
ST_CLOSE?
Returns the pressure that the switch closed at in the
current pressure units.
ST_OPEN?
Returns the pressure that the switch opened at in the
current pressure units.
ST_DEAD?
Returns deadband of the switch in the current pressure
units.
TEMP_UNIT
This command is used to set the temperature unit used
when temperature is measured.
The first argument indicates which display to apply the
change to. The second argument is the unit, either CEL
for Celsius or FAR for Fahrenheit.
For example:
To set the temperature unit to Fahrenheit on the lower
display, send TEMP_UNIT LOWER, FAR.
58
TEMP_UNIT?
Returns the temperature unit, (CEL or FAR) used when
RTDs are measured for each of the 3 displays.
VAL?
Returns the value of any measurement on the upper and
lower display. For example, if the upper display shows
5 mA, and the lower display shows 10 V, then VAL?
returns:
5.000000E-03, A, 1.000000E+01, V
ZERO_MEAS
Zeros the attached pressure module. Enter the value
zeroed in PSI after the command when an absolute
pressure module is zeroed.
ZERO_MEAS?
Returns the zero offset or the reference value for absolute
pressure modules.