Pressure calibrator
Hide thumbs Also See for 719PRO:
Table of Contents

Advertisement

Command
Returns user to manual operation of
LOCAL
the Product
Locks out the keypad of the Product in
LOCKOUT
remote operation
Sets the 0 % of span limit for percent
LO_ERR
error mode
Returns the 0 % of span limit for
LO_ERR
percent error mode
MOTOR_ON
Turns the motor on.
MOTOR_OFF Turns the motor off.
Returns the current state of the Hart
MOTOR?
resistor
Returns ohms value measured from
OHMS?
the RTD
Set the Product to output the
OUT
requested current.
Returns the value of the current being
OUT?
simulated.
Description
Command
Set the pressure unit for the indicated
PRES_UNIT
display
Returns the pressure from the
PRES_UNIT?
indicated display
Sets the approximate value at which
PUMP_LIMIT
the pump will turn off.
Returns the approximate value at
PUMP_LIMIT?
which the pump will turn off.
REMOTE
Puts the Product in remote mode
RTD_TYPE
Sets the RTD type
RTD_TYPE?
Returns the RTD type
Set the Product to simulate the
SIM
requested current.
Pressure Calibrator
Remote Operation