Command
Returns the value of the current being
SIM?
simulated.
Returns pressure value at which the
ST_CLOSE?
switch closed
Returns pressure value of the
ST_DEAD?
deadband of the switch
Returns pressure value at which the
ST_OPEN?
switch opened
ST_START
Starts a switch test
Set the RTD to read in F or C on the
TEMP_UNIT
indicated display
TEMP_UNIT?
Returns the unit the RTD is set to
VAL?
Returns the measured values
ZERO_MEAS
Zeros the pressure module
Returns the zero offset of the
ZERO_MEAS?
pressure module