Fluke 719PRO User Manual page 43

Pressure calibrator
Hide thumbs Also See for 719PRO:
Table of Contents

Advertisement

Example:
A pressure transmitter under test is 30 psi (2 Bar) Full-
Scale and regulates a corresponding 4 mA to 20 mA
signal. Program a 0 psi to 30 psi pressure span into the
Product and the Product calculates and shows the
deviation or %Error from the expected 4 mA to 20 mA
output. This eliminates the need for manual calculations
and helps when it is difficult to set an exact pressure with
an external pump.
To use the %ERROR function:
1.
From the Main Menu, push .
2.
Push .
3.
Push  to start the %ERROR option.
4.
Push  to configure the option. The first option sets
the Port.
5.
Push  to scroll through the port choices.
6.
When finished, push .
7.
Loop Power can be toggled on/off as seen below.
Push  when done.
Pressure Calibrator
Transmitter Calibration
hix049.eps
hix050.eps
33