Measure Linear Fidelity - HP ESA-L1500A User's And Calibration Manual

Spectrum analyzer
Table of Contents

Advertisement

Measure Linear Fidelity

1. Set the 1 dB and 10 dB step attenuators as indicated in
for the 4 dB from REF LVL setting.
2. Press
Peak Search
reading in column 8 of
3. Calculate the Linear Fidelity Error (LFE) as a percentage of
reference level (RL), and record the result in the performance
verification test record as indicated in column 9 of
LFE(% of RL) = 100 x (Actual Mkr Reading
/223.6 mV
For example, if the Ideal Mkr Reading is 89.3 mV and the Actual
Mkr Reading is 85 mV, the Linear Fidelity Error would be:
LFE
= 100 x (85
89.3) / 223.6
= 100 x ( 4.3) / 223.6
= 100 x 0.0192
= 1.92% of RL
4. Repeat
step 1
through
Level settings indicated in
5. Press
,
Preset
System, Alignments, Auto Align, All
6. 75
Input only: Press
Chapter 9
on the analyzer and record the Mkr amplitude
Table
9-10.
step 3
above for the remaining dB from Ref
Table
9-10.
Amplitude, More, Input Z Corr 75
Performance Verification Tests
8. Scale Fidelity
Table 9-10
Table
9-10.
Ideal Mkr Reading)
on the analyzer.
225