Procedure; Db Bandwidths - HP ESA-L1500A User's And Calibration Manual

Spectrum analyzer
Table of Contents

Advertisement

Performance Verification Tests
10. Resolution Bandwidth Accuracy
Figure 9-9
Resolution Bandwidth Accuracy Test Setup
Use only 75
CAUTION
connectors, or damage to the connectors will occur.

Procedure

1. Connect the equipment as shown in

3 dB Bandwidths

2. Set the synthesized signal generator controls as follows:
3. Press
234
cables, connectors, or adapters on instruments with 75
,
AMPLITUDE
0 dBm
,
AMPLITUDE
6 dBm (75
,
FREQUENCY
50 MHz
on the analyzer, then wait for the preset routine to
Preset
finish. Press
System, Alignments, Auto Align, Off
pressing the following keys:
,
Frequency
50 MHz
,
Span
4.5 MHz
,
BW/Avg
3 MHz
Video BW Auto Man
,
Amplitude
Scale/Div 1 dB
Figure
(50
Input only)
Input only)
,
30 Hz
9-9.
. Set the analyzer by
Chapter 9