Figure 9-7
Scale Fidelity Test Setup
Use only 75
CAUTION
connectors, or damage to the connectors will occur.
Procedure
Calculate Actual Attenuation Values
1. From the calibration data supplied with the 1 dB step attenuator,
2. From the calibration data supplied with the 10 dB step attenuator,
Chapter 9
cables, connectors, or adapters on instruments with 75
enter into column 3 of
attenuation for the corresponding nominal attenuation settings. If
the calibration data does not indicate an actual attenuation value for
the 0 dB setting, enter 0 dB. If using a programmable attenuator,
enter the data for the section three 4 dB step.
enter into column 5 of
attenuation for the corresponding nominal attenuation settings. If
Table 9-9
and
Table 9-10
Table 9-9
and
Table 9-10
Performance Verification Tests
8. Scale Fidelity
the actual
the actual
219