HP ESA-L1500A User's And Calibration Manual page 314

Spectrum analyzer
Table of Contents

Advertisement

Hewlett-Packard Company
Model HP ESA-L1500A
Serial No. ___________
Test Description
28
32
36
40
44
48
52
56
60
64
68
Linear Mode
dB from Ref Level
0
4
8
12
16
20
9.
Input Attenuation
Switching Uncertainty
Attenuation
0
5
15
Minimum
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0.4 dB
0 (Ref)
3.0%
3.0%
3.0%
3.0%
3.0%
0.30 dB
0.30 dB
0.30 dB
Chapter 10
HP ESA-L1500A Performance Verification Test Record
Report No. ___________
Date ___________
Results
Maximum
Measured
(28) ___________
(29) ___________
(30) ___________
(31) ___________
(32) ___________
(33) ___________
(34) ___________
(35) ___________
(36) ___________
(37) ___________
(38) ___________
0 (Ref)
(39) ___________
(40) ___________
(41) ___________
(42) ___________
(43) ___________
(1) ___________
+0.30 dB
(2) ___________
+0.30 dB
(3) ___________
+0.30 dB
Performance Verification Test Record
Measurement
Uncertainty
+0.4 dB
0.133 dB
+0.4 dB
0.133 dB
+0.4 dB
0.133 dB
+0.4 dB
0.132 dB
+0.4 dB
0.133 dB
+0.4 dB
0.133 dB
+0.4 dB
0.133 dB
+0.4 dB
0.149 dB
+0.4 dB
0.149 dB
+0.4 dB
0.150 dB
+0.4 dB
0.150 dB
0 (Ref)
+3.0%
1.21%
+3.0%
0.63%
+3.0%
0.40%
+3.0%
0.22%
+3.0%
0.16%
0.192 dB
0.192 dB
0.179 dB