Performance Verification Tests
24. Non-Harmonic Spurious Outputs, Options 1DN and 1DQ
24. Non-Harmonic Spurious Outputs,
Options 1DN and 1DQ
The tracking generator output is connected to the input of a microwave
spectrum analyzer. The tracking generator is set to several different
output frequencies.
For each output frequency, several sweeps are taken on the microwave
spectrum analyzer over different frequency spans and the highest
displayed spurious response is measured in each span. Responses at
the fundamental frequency of the tracking generator output or their
harmonics are ignored. The amplitude of the highest spurious response
is recorded.
There are no related adjustments for this performance test.
Equipment Required
Additional Equipment Required for 75
Procedure
1. Press
Use only 75
CAUTION
connectors or damage to the connectors will occur.
302
Spectrum analyzer, microwave
Cable, Type-N, 62-cm (24-in)
Cable, BNC, 23-cm (9-in)
Adapter, Type-N (m) to BNC (f)
Pad, minimum loss
Adapter, Type-N (f) to BNC (m), 75
on the spectrum analyzer, then wait for the preset
Preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
,
Frequency
50 MHz
,
Span
Zero Span
,
Source Amptd
Amplitude On Off
,
Source Amptd
Amplitude On Off
(75 ) RF Output only.
cables, connectors, or adapters on instruments with 75
(On),
(50
RF Output only)
0 dBm
(On),
+42.76 dBmV
RF Out
Chapter 9