The following steps apply only to analyzers with 75
NOTE
19.In
20.In
21.In
Chapter 9
Table
9-22, locate the most positive Noise Level Amplitude for
frequencies between 1 MHz and 500 MHz. Record this amplitude as
TR Entry 5 of the performance verification test record.
Table
9-22, locate the most positive Noise Level Amplitude for
frequencies greater than 500 MHz and less than or equal to
1000 MHz. Record this amplitude as TR Entry 6 of the performance
verification test record.
Table
9-22, locate the most positive Noise Level Amplitude for
frequencies greater than 1000 MHz. Record this amplitude as TR
Entry 7 of the performance verification test record.
20. Tracking Generator Feedthrough Options IDN and IDQ
Performance Verification Tests
RF Out.