5. On the analyzer, press the following keys:
6. Set the 10 dB step attenuator and analyzer reference level according
When entering reference levels on analyzers with 75
NOTE
values in dBmV with 0.01 dB precision. That is, 8.24 dBmV not
8.2 dBmV. The reference level annotation rounds to the nearest 0.1 dB,
but internal values are to the nearest 0.01 dB.
Chapter 9
Single Sweep
Peak Search
,
Marker
Marker
to
Table
9-7. At each setting, do the following:
a. Press
Single Sweep
b. Press
Peak Search
c. Record the marker delta amplitude reading in column 5 of
Table
9-7.
d. Add the actual attenuation error to the analyzer marker
amplitude and enter the result in the performance test record.
Example: for 40 dBm reference level:
Analyzer marker
Attenuation Error (50 dB) = 0.07 dB
TR Entry 4
=
0.17 dB 0.07 dB
on the analyzer.
amplitude = +0.17 dB
=
0.10 dB
Performance Verification Tests
7. Reference Level Accuracy
inputs, enter the
215