Fluke Data Acquisition System/Data Logging System 2680A Getting Started Manual page 25

Data acquisition system/data logging system
Hide thumbs Also See for Data Acquisition System/Data Logging System 2680A:
Table of Contents

Advertisement

The network number of the device must match that of the gateway. For
example, if the gateway IP address is 129.196.180.93, and the subnet mask is
255.255.255.0, the network number is 129.196.180.0, and the device IP
address must be in the range 129.196.180.1 to 129.196.180.254.
Table 7. Reviewing and Setting the Subnet Mask and Default Gateway

Press the COMM key to review the parameters, or press and hold the COMM
key for 3 seconds to set the parameters.

Press the up/down arrow keys until dgAtE (default gateway) appears

in the primary display (COMM appears in the secondary display).

Press the ENTER key. dgAtE appears in the secondary display, and ON or
OFF is in the primary display.

Press the up/down arrow keys to select either ON or OFF when in set

mode.

Press the ENTER key to make the selection (pressing any other key will cancel
set operations.). If you select ON, the subnet mask appears. The subnet mask
display consists of four 3 digit segments: Sub0. Sub1. Sub2. Sub3.

Press the left/right arrow keys to select the desired number in each segment.
The selected digit is highlighted and the segment, for example Sub:0,

appears in the secondary display.

Press the up/down arrow keys to select the desired number 0 to 9, for the
positioned subnet mask digit. In this manner, select all 12 subnet mask

digits.

Press the ENTER key to make the selection (pressing any other key will cancel
set operations). If you enter an incorrect subnet mask, Error appears for 2
seconds, and the subnet mask selection stays displayed. Otherwise, the
default gateway IP address appears. The default gateway display consists of
four 3 digit segments: gAt0.gAt1.gAt2.gAt3.

Press the left/right arrow keys to select the desired number in each segment.
The selected digit becomes highlighted and the segment, e.g. gAt:0,

appears in the secondary display.

Press the up/down arrow keys to select the desired number 0 to 9, for the
positioned default gateway IP digit. In this manner, select all 12 default

gateway IP digits.

Press the ENTER key to make the selection (pressing any other key will cancel
set operations). If you enter an incorrect default gateway IP, Error displays for 2
seconds, and the default gateway IP selection stays displayed. Otherwise, the
procedure exits.
Getting Started
Configuring the 268XA
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

2686a

Table of Contents