Configuring A Computed Channel - Fluke Data Acquisition System/Data Logging System 2680A Getting Started Manual

Data acquisition system/data logging system
Hide thumbs Also See for Data Acquisition System/Data Logging System 2680A:
Table of Contents

Advertisement

Alarm 1 or Alarm 2
Each channel has two alarms. You can set the alarms to Off, Low, or High. An
alarm condition occurs when a measurement falls below a low alarm value
(Low) or rises above a high alarm value (High). Fluke DAQ records all alarm
conditions in a data file.
Open Thermocouple Detection
Provides for notification if an open thermocouple is detected.
Factors (Mx+B)
Used to scale measurements. Mx + B scaling multiplies a measurement by a
multiplier (M) and then applies an offset (B). For example, Mx+B scaling of
100x+50 applied to a measured value of 1.15 would result in a reading of
100(1.15)+50=165.
Input or Scaled Range
Use the M and B values to calculate the Input and Scaled range or you can
enter the Input and Scaled range and Fluke DAQ will calculate the M and B
values for you.

Configuring a Computed Channel

Computed channels provide a means to calculate values based on measurements.
For example, if Channel 1 measures dc V and Channel 2 measures dc I, you could
use a computed channel to calculate instantaneous power (Channel 1 x Channel 2).
For a computed channel to execute, it must be associated with the scanning of a
module. You need to choose a module in the Scan with Module area for this
association. The default is module 1.
To configure a computed channel
1. Highlight the Computed Channels icon in the TreeView panel and click the
Insert button.
Managing Your Network Using Fluke DAQ
Getting Started
alg116s.bmp
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

2686a

Table of Contents